3 Short-Range Comms & IoT Mode
3.7 ACP Measurement
Remote
Command
[:SENSe]:ACPower:OFFSet[1]|2[:OUTer]:LIST:SIDE NEGative | BOTH | POSitive, …
[:SENSe]:ACPower:OFFSet[1]|2[:OUTer]:LIST:SIDE?
Subopcode: 1 = BTS/Downlink (Default), 2 = MS/Uplink
Example
:ACP:OFFS:LIST:SIDE BOTH
:ACP:OFFS:LIST:SIDE?
Notes Some Modes do not support Offset subopcode 2. In those cases, commands with subopcode 2 are
accepted without error but ignored
If you set POS or NEG in an offset, result of the inactive side returns -999
Preset When "Max Num of Offsets" on page 902 is 12, the preset value of Offset G ~ L is BOTH
BOTH, BOTH, BOTH, BOTH, BOTH, BOTH | BOTH, BOTH, BOTH, BOTH, BOTH, BOTH
State Saved Saved in instrument state
Range
NEGative|BOTH|POSitive
Method
Allows you to turn RRC filtering of each offset on or off. The value (roll off) for the
filter will be set to the value of the Filter Alpha parameter.
Remote
Command
[:SENSe]:ACPower:OFFSet[1]|2[:OUTer]:LIST:FILTer[:RRC][:STATe] ON | OFF | 1 |
0,…
[:SENSe]:ACPower:OFFSet[1]|2[:OUTer]:LIST:FILTer[:RRC][:STATe]?
Subopcode: 1 = BTS/Downlink (Default), 2 = MS/Uplink
Example
:ACP:OFFS:LIST:FILT 1,0,0
:ACP:OFFS:LIST:FILT?
Notes 1|ON = RRC Weighted, 0|OFF = Integ BW
Some Modes do not support Offset subopcode 2. In those cases, commands with subopcode 2 are
accepted without error but ignored
Preset When "Max Num of Offsets" on page 902 is 12, the preset value of Offset G ~ L is the same as the Offset
F value