6 Input/Output
6.3 External Gain
6.3 External Gain
Contains controls that allow you to compensate for gain or loss in the measurement
system outside the instrument. The External Gain is subtracted from the amplitude
readout (or the loss is added to the amplitude readout). So, the displayed signal
level represents the signal level at the output of the device-under-test, which can be
the input of an external device that provides gain or loss.
Entering an External Gain value does not affect the Reference Level, therefore the
trace position on screen changes, as do all of the values represented by the trace
data. Thus, the values of exported trace data, queried trace data, marker
amplitudes, trace data used in calculations such as NdB points, trace math, peak
threshold, etc., are all affected by External Gain. Changing the External Gain, even
on a trace that is not updating, immediately changes all of the above, without new
data needing to be taken.
NOTE
Changing the External Gain causes the instrument to immediately stop the
current sweep and prepare to begin a new sweep. The data will not change until
the trace data updates because the offset is applied to the data as it is taken. If a
trace is exported with a nonzero External Gain, the exported data will contain the
trace data with the offset applied.
In Spectrum Analyzer Mode, a Preamp is the common external device providing gain
or loss. In a measurement application mode like GSM or W-CDMA, the gain or loss
could be from a BTS (Base Transceiver Station) or an MS (Mobile Station). So, in the
Spectrum Analyzer mode MS and BTS would be grayed out and the only choice
would be Ext Preamp. Similarly, in some of the digital communications applications,
Ext Preamp will be grayed out and you would have a choice of MS or BTS.
The Ext Preamp, MS, and BS controls may be grayed-out depending on which
measurement is currently selected. If any of the grayed-out controls are pressed, or
the equivalent SCPI command is sent, an advisory message is generated.
6.3.1 External Preamp
This function is similar to the reference level offset function. Both affect the
displayed signal level. Ref Lvl Offset is a mathematical offset only, no instrument
configuration is affected. Ext Preamp gain is used when determining the auto-
coupled value of the Attenuator. The External Gain value and the Maximum Mixer
Level settings are both part of the automatic setting equation for the RF attenuation
setting. (10dB of Attenuation is added for every 10dB of External Gain.)
2095 Short Range Comms & IoT Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals