3 Short-Range Comms & IoT Mode
3.6 Power Stat CCDF Measurement
Normal and places it at the center of the screen on the trace determined by the
Marker Trace rules. At the same time, Marker X Axis Value appears on the Active
Function area.
The default active function is the active function for the currently selected marker
control mode. If the current control mode is OFF, there is no active function, and the
active function is turned off.
Remote
Command
:CALCulate:PSTatistic:MARKer[1]|2|…|12:MODE POSition | DELTa | OFF
:CALCulate:PSTatistic:MARKer[1]|2|…|12:MODE?
Example
:CALC:PST:MARK:MODE POS
:CALC:PST:MARK:MODE?
Preset
OFF
State Saved Saved in instrument state
Range
POSition|DELTa|OFF
Annotation Mkr # <X value> and <Marker value> upper right on graph
When Marker Trace is Polar in WCDMA mode:
Mkr # <Chip Value (RHO & QPSKEVM)/Symbol Value (CDP)>, <X value> and <Y value> upper right on
graph
Backwards Compatibility SCPI Commands
Sets or queries the state of a marker. Setting a marker that is OFF to ON (1) puts it
into Normal mode and places it at the center of the screen.
Example
:CALC:PST:MARK3:STAT 1
:CALC:PST:MARK3:STAT?
Preset
OFF
State Saved Saved in instrument state
Range
OFF|ON
Backwards
Compatibility
SCPI
:CALCulate:PSTatistic:MARKer[1]|2|…|12:STATe OFF | ON | 0 | 1
:CALCulate:PSTatistic:MARKer[1]|2|…|12:STATe?
Delta Marker (Reset Delta)
Pressing this button has exactly the same effect as pressing Delta in "Marker Mode"
on page 751. The selected marker becomes a Delta marker. If the selected marker is
already a Delta marker, the reference marker is moved to the current position of the
selected marker, thus resetting the Delta to zero.
752 Short Range Comms & IoT Mode User's &Programmer's Reference