9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
Command
:STATus:QUEStionable:INTegrity:SIGNal:ENABle?
Example
:STAT:QUES:INT:SIGN:ENAB 4
Burst Not Found will be reported to the Integrity Summary of the Status Questionable register
Preset 32767
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Questionable Integrity Signal Event Query
Returns the decimal value of the sum of the bits in the Questionable Integrity Signal
Event register.
NOTE
The register requires that the associated PTRansition or NTRansition filters be
set before a condition register bit can set a bit in the event register. The data in
this register is latched until it is queried. Once queried, the register is cleared.
Remote Command
:STATus:QUEStionable:INTegrity:SIGNal[:EVENt]?
Example
:STAT:QUES:INT:SIGN?
Preset 0
Status Bits/OPC depend-
encies
Sequential command
Questionable Integrity Signal Negative Transition
Determines which bits in the "Questionable Integrity Signal Condition" on page 2502
register will set the corresponding bit in the "Questionable Integrity Signal Event
Query" on page 2503 register when the condition register bit has a negative
transition (1 to 0).
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
Command
:STATus:QUEStionable:INTegrity:SIGNal:NTRansition <integer>
:STATus:QUEStionable:INTegrity:SIGNal:NTRansition?
Example
:STAT:QUES:INT:SIGN:NTR 4
Burst not found being cleared will be reported to the Integrity Summary of the Status Questionable
register
Preset 0
2503 Short Range Comms & IoT Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals