9 Programming the Instrument
9.2 IEEE 488.2 Common Commands
Remote
Command
*RST
Notes Sequential
Couplings *RST causes the currently running measurement to be aborted and causes the default measurement
to be active. *RST gets the mode to a consistent state, with all of the default couplings set
Status Bits/OPC
dependencies
Clears all pending OPC bits. The "Status Byte Register" on page 2459 is set to 0
9.2.10 *SAV - Save Instrument State
Saves the current instrument state and mode to the specified instrument memory
register.
Remote
Command
*SAV <register #>
Example Save the instrument state in register 9 (register 10 in the UI):
*SAV 9
Notes Registers 0 through 15 are accessible from the front panel in menu keys for Save Registers. Register 0
corresponds to the front panel Register 1
Min/Max 0 / 127
Status Bits/OPC
dependencies
The command is sequential
9.2.11 *SRE - Service Request Enable
Enables the desired bits of the "Service Request Enable Register" on page 2462.
The query returns the value of the register, indicating which bits are currently
enabled.
Numeric values for bit patterns can be entered using decimal or hexadecimal
representations (0 to 32767, equivalent to #H0 to #H7FFF).
Remote
Command
*SRE <integer>
*SRE?
Example Enable bits 1, 2, and 4 in the service request enable register:
*SRE 22
Notes For related commands, see "Status Register System & STATus Subsystem" on page 2450 and
:SYSTem:ERRor[:NEXT]?
Preset 0
Min/Max 0 / 255
2427 Short Range Comms & IoT Mode User's &Programmer's Reference