9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
Questionable Calibration Skipped Enable
Determines which bits in the "Questionable Calibration Skipped Condition" on page
2495 Register will set bits in the "Questionable Calibration Skipped Event Query" on
page 2496 register, which also sets bit 11 of the "Questionable Calibration Register"
on page 2484.
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
Command
:STATus:QUEStionable:CALibration:SKIPped:ENABle <integer>
:STATus:QUEStionable:CALibration:SKIPped:ENABle?
Example
:STAT:QUES:CAL:SKIP:ENAB 1
Can be used to query if an RF alignment skipped condition is detected
Preset 32767
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Questionable Calibration Skipped Event Query
Returns the decimal value of the sum of the bits in the Questionable Calibration
Event register.
NOTE
The register requires that the associated PTRansition or NTRansition filters be
set before a condition register bit can set a bit in the event register. The data in
this register is latched until it is queried. Once queried, the register is cleared.
Remote Command
:STATus:QUEStionable:CALibration:SKIPped[:EVENt]?
Example
:STAT:QUES:CAL:SKIP?
Preset 0
Status Bits/OPC depend-
encies
Sequential command
Questionable Calibration Skipped Negative Transition
Determines which bits in the "Questionable Calibration Skipped Condition" on page
2495 register will set the corresponding bit in the "Questionable Calibration Skipped
Short Range Comms & IoT Mode User's &Programmer's Reference 2496