9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
Extended Needed Event Query" on page 2490 register when the condition register
bit has a positive transition (0 to 1).
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
Command
:STATus:QUEStionable:CALibration:EXTended:NEEDed:PTRansition <integer>
:STATus:QUEStionable:CALibration:EXTended:NEEDed:PTRansition?
Example
:STAT:QUES:CAL:EXT:NEED:PTR 2
Conducted alignment required bit being set will be reported
Preset 32767
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
9.4.6.11 Questionable Calibration Extended Failure Register
Monitors conditions that occur because a calibration or alignment has failed to
complete properly. It summarizes them in bit 9 of the "Questionable Calibration
Register" on page 2484.
Bit Condition Operation
2 Characterize Preselector
Failure
The preselector characterization failed
5 Align Selected Freq Range
failed
The alignment for selected frequency range failed
8 MPAdapter Preamp Charact
Failure
The Multiport Adaptor must be calibrated (EXT only)
9 RCal Apply Warning, Check
RCal Status
The calibration request sent to the RCal module failed
2491 Short Range Comms & IoT Mode User's &Programmer's Reference