3 Short-Range Comms & IoT Mode
3.4 HRP UWB Demodulation Measurement
Remote Command
[:SENSe]:HUWB:AVERage:COUNt <integer>
[:SENSe]:HUWB:AVERage:COUNt?
Example
:HUWB:AVER:COUN 1000
:HUWB:AVER:COUN?
Preset 10
State Saved Yes
Min 1
Max 10000
Averaging On/Off
Turns averaging on or off.
Remote Command
[:SENSe]:HUWB:AVERage[:STATe] OFF | ON | 0 | 1
[:SENSe]:HUWB:AVERage[:STATe]?
Example
:HUWB:AVER OFF
:HUWB:AVER?
Preset
OFF
State Saved Yes
Range
OFF|ON
Average Mode
Toggles the averaging mode. This selection only affects the averaging result after
the number of N averages is reached. The value of N is set using "Avg | Hold
Number" on page 497.
EXPonential
Each successive data acquisition after the average count is reached is
exponentially weighted and then combined with the existing average
REPeat
After reaching the average count, averaging is reset, and a new average is
started
Remote
Command
[:SENSe]:HUWB:AVERage:TCONtrol EXPonential | REPeat
[:SENSe]:HUWB:AVERage:TCONtrol?
Example
:HUWB:AVER:TCON REP
:HUWB:AVER:TCON?
Notes Selects the type of termination control used for averaging. This determines the averaging action after
the specified number of frames (average count) is reached
Preset
EXPonential
498 Short Range Comms & IoT Mode User's &Programmer's Reference