5 Preset
5.7 Save User Preset
5.7 Save User Preset
Saves the state of the currently active Mode in a unique location, for recall by the
key "User Preset" on page 1907. Each Mode has one such location, so, for each
Mode, one User Preset can be defined.
Save User Preset can be executed from the "Preset Dropdown" on page 1901.
All the Mode variables are saved, including those reset by "Mode Preset" on page
1902, those only reset by "Restore Mode Defaults" on page 1904, and all
Input/Output variables, so when you subsequently press User Preset, the instrument
returns to the exact same setup that existed when you pressed Save User Preset.
Thus, User Preset has wider scope than Mode Preset.
Remote
Command
:SYSTem:PRESet:USER:SAVE
Example
:SYST:PRES:USER:SAVE
Notes :SYST:PRES:SAVE creates the same file as if you requested *SAV or :MMEM: STOR:STAT,
except that Save User Preset does not allow you to specify the file name or location
1909 Short Range Comms & IoT Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)