3 Short-Range Comms & IoT Mode
3.10 Spurious Emissions Measurement
View Size Position
Graph + Metrics Two fifth, full width Bottom
Gate View One third, full width Bottom
Measure Trace
See "Measure Trace" on page 1508.
Trace Type
This is the trace type (and view/blank parameter) of a trace specified by Measure
Trace.
3.10.2.3 All Range Table
Window #4
The spurs listed are within the current value of the Marker Peak Excursion setting of
the absolute limit. All the spurs listed passed. Any spur that has failed the absolute
limit displays ‘F’ beside it.
Result Units Min Max
Spur N/A 0 200
Range N/A 1 20
Start Freq See "Start Freq" on page 1433 under Meas Setup
Stop Freq See "Stop Freq" on page 1434 under Meas Setup
RBW See "Res BW" on page 1440 under Meas Setup
Frequency Hz Analyzer Min Analyzer Max
Amplitude dBm Analyzer Min Analyzer Max
Limit dBm -200 50
Δ Limit dBm (Limit - Amplitude)
Views in which the All Range Table window appears:
View Size Position
All Ranges Two fifth, full width Bottom
Gate View One third, full width Bottom
1360 Short Range Comms & IoT Mode User's &Programmer's Reference