7 Save/Recall/Print
7.3 Save
Both single and double quotes are supported for any filename parameter over remote
After saving to a register, that register’s menu key is updated with the date and time of the save
After saving to a register, you remain in the Save Trace menu, so that you can see the Register key
update. After saving to a file, the instrument automatically returns to the previous menu and any Save
As dialog goes away
More Information
In measurements that support saving Traces, for example, Swept SA, the Trace data
is saved along with the State in the State file. When recalling the State, the Trace
data is recalled as well. Traces are recalled exactly as they were stored, including
the writing mode and update and display modes. If a Trace was updating and visible
when the State was saved, it returns updating and visible, and its data will be
rewritten right away. When you use State to save and recall traces, any trace whose
data must be preserved should be placed in View or Blank mode before saving.
The following table describes the Trace Save and Recall possibilities:
You want to
recall state and
one trace’s
data, leaving
other traces
unaffected
Save Trace+State from 1 trace.
Make sure that no other traces
are updating (they should all be
in View or Blank mode) when
the save is performed
On recall, specify the trace you want to load
the one trace’s data into. This trace loads in
view. All other traces’ data will be unaffected,
although their trace mode will be as it was
when the state save was performed
You want to
recall all traces
Save Trace+State from ALL
traces
On recall, all traces come back in View (or
Blank if they were in Blank or Background
when saved)
You want all
traces to load
exactly as they
were when
saved
Save State On recall, all traces’ mode and data will be
exactly as they were when saved. Any traces
that were updating will have their data
immediately overwritten
7.3.6 Save From Trace
Selects the trace to be saved. The default is the currently selected trace, selected in
this this or any other menu with Trace selection. If you have chosen All then it
remains chosen until you specifically change it to a single trace, regardless of the
trace selected in the Trace menu.
When you select a trace, it makes that trace the current trace, so it displays on top
of all of the other traces.
2262 Short Range Comms & IoT Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals