6 Input/Output
6.1 RF Source
Steps" on page 1934. As a step is setup, the value entered runs through several
levels of validation.
Remote Command
:SOURce:LIST:SETup:INPut:TRIGger <enum>, <enum>, <enum>, …
:SOURce:LIST:SETup:INPut:TRIGger?
Example
:SOUR:LIST:SET:INP:TRIG IMM,INT,EXT2
:SOUR:LIST:SET:INP:TRIG?
Notes The command is to setup below parameter array of whole list sequence
Step Trigger <enum> - specifies the input trigger for the step. For details of the valid types of step
trigger see "Step Trigger" on page 1935
If input parameter number exceeds the step number defined by "Number of Steps" on page 1934,
then error -221 "Settings conflict; The number of input parameters is too large and is truncated to
current list step number" is generated, and only those parameters whose index number falls in
number of steps will be updated
Dependencies The range is 1 to 1000, which is determined by the number of steps you have configured. For details
see "Number of Steps" on page 1934
Step Configuration of Transition Time parameter list (Remote Command Only)
Configures the “Transition Time” parameter array of the whole List Sequencer at
once. The array size is the same as step number defined in "Number of Steps" on
page 1934. As a step is setup, the value entered runs through several levels of
validation.
Remote Command
:SOURce:LIST:SETup:TRANsition:TIME <time>, <time>, <time>, …
:SOURce:LIST:SETup:TRANsition:TIME?
Example
:SOUR:LIST:SET:TRAN:TIME 1ms,1ms,1ms
:SOUR:LIST:SET:TRAN:TIME?
Notes The command is to setup below parameter array of whole list sequence
Transition Time <time> - specifies the transition time for the step in seconds. For details of the valid
ranges for the transition time see "Transition Time" on page 1936
If input parameter number exceeds the step number defined by "Number of Steps" on page 1934,
then the error -221, "Settings conflict; The number of input parameters is too large and is truncated to
current list step number" is generated, and only those parameters whose index number falls in
number of steps will be updated
Dependencies The range is 1 to 1000 which is determined by the number of steps you have configured. For details
see "Number of Steps" on page 1934
Step Configuration of Radio Band parameter list (Remote Command Only)
Configures the Radio Band parameter array of the whole List Sequencer at once.
The size of the array is the same as the step number defined in "Number of Steps"
Short Range Comms & IoT Mode User's &Programmer's Reference 1950