3 Short-Range Comms & IoT Mode
3.8 SEM Measurement
Remote Command
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:SWEep:TYPE SWEep | FFT, …
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:SWEep:TYPE?
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:SWEep:TYPE:AUTO ON | OFF | 1 | 0, …
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:SWEep:TYPE:AUTO?
Example
:SEM:OFFS2:INN:LIST:SWE:TYPE FFT,FFT,SWE
:SEM:OFFS2:INN:LIST:SWE:TYPE?
:SEM:OFFS2:INN:LIST:SWE:TYPE:AUTO ON, ON, ON, ON, OFF, OFF
:SEM:OFFS2:INN:LIST:SWE:TYPE:AUTO?
Notes Comma-separated list of values
OFFSet 1 is for BTS, 2 for MS. Default is BTS
Dependencies Not available in modular products, such as VXT
Couplings When Sweep Type is set manually, Sweep Type Mode is set to MANual
When Sweep Type Mode is Auto, Sweep Type is automatically selected according to "Sweep Type
Rules" on page 1097
Preset Automatically calculated
ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON
State Saved Saved in instrument state
Range Auto|Man
Offset Side
Specifies which offset side to measure.
You can turn off (not use) specific offsets with [:SENSe]:SEMask:OFFSet
[n]:INNer:LIST:STATe.
BOTH
Both sides in the sub-block gap are enabled.
NEGative
The upper side in the sub-block gap only (i.e., negative sideband of the upper sub-
block) is enabled
POSitive
The lower side in the sub-block gap only (i.e., positive sideband of the lower sub-
block) is enabled.
When sending the remote command, missing values are not permitted; that is, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values
remain unchanged. The query for this parameter always returns 12 values.
The figure below shows the relation between the negative/positive offset side
setups and the upper/lower results in the MSR and LTE-Advanced FDD/TDD
Modes.
Short Range Comms & IoT Mode User's &Programmer's Reference 1151