3 Short-Range Comms & IoT Mode
3.8 SEM Measurement
Rules
Preset Automatically calculated
LTEAFDD, LTEATDD, 5G NR, MSR Modes:
ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON|
ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON
When the max number of offsets is 14:
WLAN Mode:
ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON
All Other Modes:
When the max number of offsets is 6:
ON,ON,ON,ON,ON,ON
When the max number of offsets is 12:
ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON
State Saved Saved in instrument state
Range Auto|Man
Offset Side
Specifies which offset side to measure.
You can turn off (not use) specific offsets with [:SENSe]:SEMask:OFFSet[n]
[:OUTer]:LIST:STATe.
BOTH
Both of the negative (lower) and positive (upper) sidebands
NEGative
Negative (lower) sideband only
POSitive
Positive (upper) sideband only
When sending the remote command, missing values are not permitted; that is, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values
remain unchanged. The query returns 14 values for WLAN Mode, and 12 values for
other Modes.
The figure below shows the relation between the negative/positive offset side
setups and the upper/lower results in the MSR, LTE-Advanced FDD/TDD and 5G NR
Modes.
Short Range Comms & IoT Mode User's &Programmer's Reference 1121