3 Short-Range Comms & IoT Mode
3.7 ACP Measurement
Avg | Hold Number
Specifies the number of measurement averages used to calculate the measurement
result. The average will be displayed at the end of each sweep. After the specified
number of average counts, the average mode (termination control) setting
determines the average action.
Remote
Command
[:SENSe]:ACPower:AVERage:COUNt <integer>
[:SENSe]:ACPower:AVERage:COUNt?
Example
:ACP:AVER:COUN 250
:ACP:AVER:COUN?
Notes The BAF backwards Compatibility SCPI command, [:SENSe]:ACPR:AVERage[:STATe], is
provided to support same functionality as [:SENSe]:ACPr:AVERage[:STATe] (PSA W-CDMA,
PSA cdma2000 and PSA 1xEVDO) due to ACPr node conflicts with ACPower node
Preset 10
State Saved Saved in instrument state
Min/Max 1/1000
Annotation Avg Number is displayed in the Measurement Bar
Backwards
Compatibility
SCPI
[:SENSe]:ACPR:AVERage:COUNt
[:SENSe]:MCPower:AVERage:COUNt
Power Suite, W-CDMA
Averaging On/Off
Turns averaging on or off.
NOTE
In this measurement, the Average Type is preset to the Log-Pwr Avg (Video)
method. Other averaging methods are not available.
Remote Command
[:SENSe]:ACPower:AVERage[:STATe] OFF | ON | 0 | 1
[:SENSe]:ACPower:AVERage[:STATe]?
Example
:ACP:AVER OFF
:ACP:AVER?
Preset
ON
State Saved Yes
Range
OFF|ON
Backwards Compatibility
SCPI
[:SENSe]:ACPR:AVERage[:STATe]
[:SENSe]:MCPower:AVERage[:STATe]
Power Suite, W-CDMA