3 Short-Range Comms & IoT Mode
3.9 Occupied BW Measurement
Remote
Command
[:SENSe]:OBWidth:AVERage[:STATe] ON | OFF | 1 | 0
[:SENSe]:OBWidth:AVERage[:STATe]?
Example
:OBW:AVER ON
:OBW:AVER?
Couplings Averaging state is coupled to "Max Hold (Remote Command Only)" on page 1299. If Max Hold is
changed from OFF to ON, Averaging state is automatically set to ON
Preset
ON
State Saved Yes
Range
ON|OFF
Backwards
Compatibility
SCPI
[:SENSe]:EBWidth:AVERage[:STATe]
Average Mode
Lets you set the averaging mode.
EXPonential
Measurement averaging continues using the specified number of averages to
compute each averaged value. The average is displayed at the end of each
sweep
REPeat
The measurement resets the average counter each time the specified number
of averages is reached
Remote
Command
[:SENSe]:OBWidth:AVERage:TCONtrol EXPonential | REPeat
[:SENSe]:OBWidth:AVERage:TCONtrol?
Example
:OBW:AVER:TCON REP
:OBW:AVER:TCON?
Preset
EXP
State Saved Yes
Range
EXPonential|REPeat
% of OBW Power
Assigns the percentage of the total power that is measured within the Occupied
Bandwidth for the current measurement. The resulting Occupied Bandwidth limits
are displayed by markers placed on the frequencies of the specified percentage.
Remote Command
[:SENSe]:OBWidth:PERCent <real>
[:SENSe]:OBWidth:PERCent?
Example
:OBW:PERC 75
1294 Short Range Comms & IoT Mode User's &Programmer's Reference