3 Short-Range Comms & IoT Mode
3.6 Power Stat CCDF Measurement
Executing Auto Coupledoes not affect markers, marker functions, trace or display
attributes, or any other instrument setting other than those specifically mentioned
above.
Measurement-Specific Details
TOI (SA Mode only)
Parameters affected by Auto Couple are:
Center Frequency Step
Resolution Bandwidth
Span/RBW Ratio
Sweep Time
Video BANDwidth VBW/RBW ratio
Upper and Lower Tone (set to Sense)
Zero span measurement Resolution Bandwidth
Zero span measurement Dwell Time
Harmonics (SA Mode only)
Parameters affected by Auto Couple are:
Resolution Bandwidth
Fundamental Frequency
Dwell Time
Range Table Resolution Bandwidths
Range Table Dwell Times
Meas Preset
Restores all the measurement parameters to their default values.
Remote
Command
:CONFigure:PSTatistic
758 Short Range Comms & IoT Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals