3 Short-Range Comms & IoT Mode
3.6 Power Stat CCDF Measurement
Executing Auto Coupledoes not affect markers, marker functions, trace or display
attributes, or any other instrument setting other than those specifically mentioned
above.
Measurement-Specific Details
TOI (SA Mode only)
Parameters affected by Auto Couple are:
–
Center Frequency Step
–
Resolution Bandwidth
–
Span/RBW Ratio
–
Sweep Time
–
Video BANDwidth VBW/RBW ratio
–
Upper and Lower Tone (set to Sense)
–
Zero span measurement Resolution Bandwidth
–
Zero span measurement Dwell Time
Harmonics (SA Mode only)
Parameters affected by Auto Couple are:
–
Resolution Bandwidth
–
Fundamental Frequency
–
Dwell Time
–
Range Table Resolution Bandwidths
–
Range Table Dwell Times
Meas Preset
Restores all the measurement parameters to their default values.
Remote
Command
:CONFigure:PSTatistic
758 Short Range Comms & IoT Mode User's &Programmer's Reference