3 Short-Range Comms & IoT Mode
3.3 LoRa (CSS) Demodulation Measurement
empirical data symbol length. The product will be further multiplied with an empirical coefficient and
eventually set as the minimum measure time
Preset 72 ms
State Saved Saved in instrument state
Min/Max 1 us / 100 s
Meas Time Auto
Remote Command
[:SENSe]:LORA:DEMod:TIME:AUTO OFF | ON | 0 | 1
[:SENSe]:LORA:DEMod:TIME:AUTO?
Example
:LORA:DEM:TIME:AUTO OFF
:LORA:DEM:TIME:AUTO?
Preset ON
Sweep/Measure
Allows users to toggle the sweep mode. The single/continuous state is meas global
which affects all measurements.
The Single/Cont key on the front panel performs this exact same function.
See "Windows" on page 315"Windows" on page 315"More Information" on page
422
Remote
Command
:INITiate:CONTinuous OFF | ON | 0 | 1
:INITiate:CONTinuous?
Example
:INIT:CONT 0
puts analyzer in Single measurement operation
:INIT:CONT OFF
puts analyzer in Single measurement operation
:INIT:CONT 1
puts analyzer in Continuous measurement operation
:INIT:CONT ON
puts analyzer in Continuous measurement operation
Preset
ON
(Note that SYST:PRESet sets INIT:CONT to ON but *RST sets INIT:CONT to OFF)
State Saved Saved in instrument state
Annunciation The single/continuous icon on the meas bar changes depending on the setting. A line with an arrow is
single, a loop with an arrow is continuous
Short Range Comms & IoT Mode User's &Programmer's Reference 421