3 Short-Range Comms & IoT Mode
3.10 Spurious Emissions Measurement
If the display is OFF, many SCPI commands related to User Views and Multiscreen
functionality do not work, and return the error messages “-221, Settings conflict;
Screen SCPI cannot be used when Display is disabled” or “221, Settings conflict;
View SCPI cannot be used while Display is disabled”. These commands include:
Name Command
Select User View
:DISPlay:VIEW:ADVanced:SELect
Rename User View
:DISPlay:VIEW:ADVanced:REName
Delete User View
:DISPlay:VIEW:ADVanced:DELete
Create User View
:DISPlay:VIEW:ADVanced:NAME
Select Screen
:INSTrument:SCReen:SELect
Delete Screen
:INSTrument:SCReen:DELete
Delete All But This Screen
:INSTrument:SCReen:DELete:ALL
Add Screen
:INSTrument:SCReen:CREate
Rename Screen
:INSTrument:SCReen:REName
Sequencer On/Off
:SYSTem:SEQuencer
Remote
Command
:DISPlay:ENABle OFF | ON | 0 | 1
:DISPlay:ENABle?
Example
:DISP:ENAB OFF
Couplings :DISP:ENAB OFF turns Backlight OFF and :DISP:ENAB ON turns Backlight ON, but changing
Backlight settings does not change the state of :DISP:ENAB
Preset
ON
Set by :SYST:DEF MISC, but not affected by *RST or :SYSTem:PRESet
State Saved Not saved in instrument state
Backwards
Compatibility
Notes
:SYST:PRES no longer turns on :DISPlay:ENABle as it did in legacy analyzers
3.10.5.3 View
Contains controls for selecting the current View, and for editing User Views.
Views
This measurement has two predefined views:
# Name SCPI
1 "Graph + Metrics" on page 1407
RESult
2 "All Ranges" on page 1407
ALL