4 System
4.5 Restore Defaults
4.5 Restore Defaults
Provides initialization of system setting groups, including the option to set the entire
instrument back to a factory default state.
NOTE
In products that run multiple instances of the X-Series Application, all instances
have the same factory default states for Restore Defaults.
Remote
Command
:SYSTem:DEFault [ALL] | ALIGn | INPut | MISC | MODes | PON | UINTerface |
SCReen
Example
:SYST:DEF
State Saved No
4.5.1 Input/Output
Input/Output Preset resets the group of settings and data associated with the
Input/Output front-panel key to their default values. These settings are not affected
by a Mode Preset because they are associated with connections to the instrument,
which you will probably not want to reset every time you press Mode Preset.
By using Input/Output Preset and "Restore Mode Defaults" on page 1904, a full
preset of the current mode will be performed, with the caveat that since
Input/Output Preset is a global function, it will affect all modes.
This is the same as the Input/Output Preset button in the Preset dropdown and the
Input/Output menu.
When Input/Output is selected, a message appears saying:
This will reset all of the Input/Output variables to their default state,
including which input is selected, all Amplitude Correction settings and data,
all External Mixing settings, all Frequency Reference settings and all Output
settings
It will not affect Alignment data or settings
This action cannot be undone. Do you want to proceed?
The dialog includes OK and Cancel controls, for you to confirm or cancel the
operation.
Example
:SYST:DEF INP
Short Range Comms & IoT Mode User's &Programmer's Reference 1796

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals