3 Short-Range Comms & IoT Mode
3.5 Channel Power Measurement
Power Limit Fail (Remote Query Only)
Queries whether a power test passes or fails. When DVB-T radio standard is
selected in SA Mode, the result of this query has no meaning.
While implementing the scpi, don’t try to remove the SCPI command from the SCPI
tree when DVB-T is selected as current radio standard.
Remote
Command
:CALCulate:CHPower:LIMit:POWer:FAIL?
Example
:CALC:CHP:LIM:POW:FAIL?
Notes Query only
When "Power Limit" on page 636 state is OFF, the returned value is always 0 (pass)
When Power Limit state is ON, the returned value is 0 (pass) if power test passes and 1(fail) if power test
fails
PSD Limit Fail (Remote Query only)
Queries whether PSD test passes or fails. When DVB-T radio standard is selected in
SA Mode, the result of this query has no meaning.
While implementing the scpi, don’t try to remove the SCPI command from the SCPI
tree when DVB-T is selected as current radio standard.
Remote
Command
:CALCulate:CHPower:LIMit:PSD:FAIL?
Example
:CALC:CHP:LIM:PSD:FAIL?
Notes Query only
When "PSD Limit" on page 637 state is OFF, the returned value is always 0 (pass)
When PSD Limit state is ON, the returned value is 0 (pass) if PSD test passes and 1(fail) if PSD test fails
3.5.8.5 Advanced
Contains controls for setting advanced functions of the instrument.
Does not appear in VXT.
Phase Noise Optimization
Lets you select the LO (local oscillator) phase noise behavior for various operating
conditions. When in Auto, selects the LO (local oscillator) phase noise behavior to
optimize dynamic range and speed for various instrument operating conditions.