3 Short-Range Comms & IoT Mode
3.8 SEM Measurement
(Primary Fail Mask
selection) OR Abs2
OR
Checks against both Primary and Abs2 limits. The test fails
if either of the limits is broken
(Primary Fail Mask
selection) AND Abs2
AND
Checks against both Primary and Abs2 limits. The test fails
if both of the limits are broken
Abs2 Disabled
OFF
Fail Mask2 is disabled
For examples, see "Fail Mask2" on page 1137.
Note that the Primary Fail Mask selection is set by "Fail Mask" on page 1156.
You can turn off (not use) specific inner offset channels remotely using:
[:SENSe]:SEMask:OFFSet[n]:INNer:LIST:STATe
When sending the command, missing values are not permitted; that is, if you want
to change values 2 and 6, you must send all values up to 6. Subsequent values
remain unchanged. The query always returns 12 values.
Remote
Command
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:TEST:SABSolute AND | OR | OFF, …
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:TEST:SABSolute?
Example
:SEM:OFFS:INN:LIST:TEST:SABS AND, AND, OR, OFF, OFF, OFF
:SEM:OFFS:INN:LIST:TEST:SABS?
Notes Comma-separated list of values
Preset
OFF, OFF, OFF, OFF, OFF, OFF, OFF, OFF, OFF, OFF, OFF, OFF | OFF, OFF, OFF,
OFF, OFF, OFF, OFF, OFF, OFF, OFF, OFF, OFF
State Saved Saved in instrument state
Range OR Abs2 | AND Abs2 | Abs2 Disabled
Meas Setup Summary Table
Lets you view and access many of the parameters in the Meas Setup menus on one
screen.
Auto Couple
Immediately puts all Auto/Man functions into Auto. Auto Couple is confined to the
current measurement only. It does not affect other measurements in the Mode.
In the Auto state, Auto/Man functions are said to be “coupled”, meaning their values
change as you make changes to other values in the measurement. This helps ensure
accurate measurements and optimum dynamic range. Auto Couple is an immediate
action function, and when it is executed, all the Auto/Man controls for the current
measurement are set to Auto, and all measurement settings coupled to the
Auto/Man parameters are automatically set to their optimal values.
Short Range Comms & IoT Mode User's &Programmer's Reference 1159