6 Input/Output
6.9 Calibration
If there are multiple Cals being executed in a Cal Group, this dialog advises you
when each Cal is complete. It also provides the ability to abort the Execute Cal
Request.If you choose to abort, calibrations that have completed use the new Cal
data and update the Last Cal field. Calibrations that have not completed retain the
existing Cal data and Last Cal timestamp, or show “---” if the Cal had never been
executed.
Remote Command
:SYSTem:CALibration:INITiate:SELected
Example
:SYST:CAL:INIT:SEL
Notes Cals cannot be applied until they have been calibrated. Once a Cal has been calibrated, the Last Cal
field in the table displays the date and time the Cal was last calibrated
Dependencies Applied to the currently selected Cal Group
Couplings Calibrate Selected is disabled if there are no Calibrate checkboxes checked. If the disabled control is
selected, the advisory message “Check the Calibrate box for the Cals you want to calibrate” is
displayed
6.9.1.3 Apply Cal Group
This is the same as "Apply Cal Group" on page 2213 in the Calibration tab.
6.9.1.4 Abort Calibration
Aborts the Calibration routine of the currently selected Cal Group
Remote Command
:SYSTem:CALibration:ABORt
Example
:SYST:CAL:ABOR
Dependencies Aborts the currently running calibration. The previously-run calibrations will still be available, but the
current calibration is halted, and next calibrations selected are not executed. Once the calibration
starts, the modal dialog appears, and the abort can be executed by selecting Cancel
2181 Short Range Comms & IoT Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals