9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
Event Query" on page 2496 register when the condition register bit has a negative
transition (1 to 0).
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
Command
:STATus:QUEStionable:CALibration:SKIPped:NTRansition <integer>
:STATus:QUEStionable:CALibration:SKIPped:NTRansition?
Example
:STAT:QUES:CAL:SKIP:NTR 1
RF Align Skipped bit being cleared will be reported
Preset 0
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Questionable Calibration Skipped Positive Transition
Determines which bits in the "Questionable Calibration Skipped Condition" on page
2495 register will set the corresponding bit in the "Questionable Calibration Skipped
Event Query" on page 2496 register when the condition register bit has a positive
transition (0 to 1).
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
Command
:STATus:QUEStionable:CALibration:SKIPped:PTRansition <integer>
:STATus:QUEStionable:CALibration:SKIPped:PTRansition?
Example
:STAT:QUES:CAL:SKIP:PTR 1
RF Align Skipped bit being set will be reported
Preset 32767
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
9.4.6.13 Questionable Integrity Register
Monitors measurement integrity-related conditions within the instrument and
summarizes them in bit 9 of the "Questionable Register" on page 2472. Two of the
bits are summary bits from lower-level event registers.
2497 Short Range Comms & IoT Mode User's &Programmer's Reference