6 Input/Output
6.5 Corrections
6.5.7.4 Correction On
Checking or unchecking this box turns the Selected Complex Correction ON or OFF.
Turning it ON causes the values in it to be applied to the data. This state transition
also automatically turns on "Apply Corrections" (sets it to ON), otherwise the
correction would not take effect.
A new sweep/acquisition is initiated if a complex correction is switched on or off.
Note that changing, sending or loading corrections data does not directly initiate a
sweep, however in general these operations will turn corrections on, which does
initiate a sweep.
Remote Command
[:SENSe]:CCORrection:CSET[:STATe] ON | OFF | 1 | 0
[:SENSe]:CCORrection:CSET[:STATe]?
Example Select correction 3:
:CCOR:CSET:SEL 3
Turn correction 3 on:
:CCOR:CSET ON
Dependencies Changing this from OFF to ON automatically turns on "Apply Corrections"
Grayed-out if Complex Corrections is not supported by the current measurement. A warning or SCPI
error is generated if you try to turn it on under these circumstances: “Feature not supported for this
measurement”
Preset Not affected by Preset. Set to OFF by Restore Input/Output Defaults
State Saved Saved in instrument state
Annotation If any Complex Correction is turned on, CC in the Meas Bar will display in amber to indicate Complex
Corrections are in use
6.5.7.5 Correction Port
Maps one of the sets of corrections to a particular I/O port. This control allows any
Input port (including External Mixing, BBIQ, the RF2 input, etc.) to be mapped to a
specific Correction, so that the Correction is only applied when that Port is being
used by the current Screen. You can also map any internal source Output port to a
specific Correction.
When Current Input (CINPut) is selected for Correction Port, it chooses the current
input port of the current Screen for the selected Correction. In other words, the
Correction applies to whichever input is selected. If the input changes, the
correction applies to the new input.
When using the VXT M9410A/11A with Remote Radio Heads (such as the Keysight
M1740A mmWave Transceiver for 5G), the choices in the dropdown menu will
appear as
Head h RFHD p
Short Range Comms & IoT Mode User's &Programmer's Reference 2128

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals