4 System
4.2 I/O Config
4.2.11.2 Config IPV4 Address (Remote Command Only)
Remote
Command
:SYSTem:COMMunicate:LAN:IPV4:CONFig <ipaddress>
:SYSTem:COMMunicate:LAN:IPV4:CONFig?
Example
:SYSTem:COMMunicate:LAN:IPV4:CONFig "192.168.1.146"
:SYSTem:COMMunicate:LAN:IPV4:CONFig?
Notes Applies to instruments that have multiple Network Adapters. When more than one network adapter is
present in the system, you must specify in the instrument config file the IP address to use to enable
Remoting channel bindings. If this is not provided, Remoting connections are likely to fail on systems
where multiple NICs are enabled
Sets the valid IPV4 address, passed in as string in the config file
The query returns IPV4 address, as a string
If config file is missing, “” (empty string) is returned
Changing the IPV4 value requires a restart of the instrument software, to ensure that servers use the
configured IP address
State Saved No
4.2.11.3 Config IPV6 Address (Remote Command Only)
Remote
Command
:SYSTem:COMMunicate:LAN:IPV6:CONFig <ipaddress>
:SYSTem:COMMunicate:LAN:IPV6:CONFig?
Example
:SYSTem:COMMunicate:LAN:IPV6:CONFig "2001:0db8:85a3:0000:0000:8a2e:0370:7334"
:SYSTem:COMMunicate:LAN:IPV6:CONFig?
Notes Applies to instruments that have multiple Network Adapters. When more than one network adapter is
present in the system, you must specify in the instrument config file the IP address to use to enable
Remoting channel bindings. If this is not provided, Remoting connections are likely to fail on systems
where multiple NICs are enabled
Sets the valid IPV6 address, passed in as string in the config file
The query returns IPV6 address, as a string
If config file is missing, “” (empty string) is returned
Changing the IPV6 value requires a restart of the instrument software, to ensure servers use the
configured IP address
State Saved No
4.2.11.4 List All Physical Network Adapter IP Addresses (Remote Query Only)
Remote
Command
:SYSTem:COMMunicate:LAN:PHYSical:IPADdress:LIST?
Example
:SYSTem:COMMunicate:LAN:PHYSical:IPADdress:LIST?
Short Range Comms & IoT Mode User's &Programmer's Reference 1768

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals