6 Input/Output
6.5 Corrections
6.5.7.1 Go To Row (Select Correction)
Specifies the selected complex correction. The selected correction will be identified
by the blue outlined row in the dialog.
The "selected complex correction" is an important concept when sending SCPI
commands to the Complex Corrections system, because in each case the SCPI
command is directed to the currently selected Complex Correction and that will be
the Correction which is modified by the SCPI command.
Remote
Command
[:SENSe]:CCORrection:CSET:SELect <integer>
[:SENSe]:CCORrection:CSET:SELect?
Example
:CCOR:CSET:SEL 3
:CCOR:CSET:SEL?
Notes The selected correction is remembered even when not in the correction menu
Preset Set to Correction 1 by Restore Input/Output Defaults
Min 1
Max 64
6.5.7.2 Delete Row
Deletes the currently-selected Complex Correction and clears all entries in that row
to the default.
Remote
Command
[:SENSe]:CCORrection:CSET:DELete
Example Select correction 3:
:CCOR:CSET:SEL 3
Delete correction 3:
:CCOR:CSET:DEL
6.5.7.3 Delete All
Deletes all complex corrections and clears all entries in all rows to the default.
When this key is pressed a prompt is placed on the screen that says “Please press
Enter or OK key to delete all complex corrections. Press ESC or Cancel to close this
dialog.” The deletion is only performed if you press OK or Enter.
Remote
Command
[:SENSe]:CCORrection:CSET:ALL:DELete
Example
:CCOR:CSET:ALL:DEL
2127 Short Range Comms & IoT Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals