5 Preset
5.15 Restart Application (Application Shutdown)
5.15 Restart Application (Application Shutdown)
Restarts the instrument application without rebooting the instrument. Before you
send this command, make sure you have saved any trace or measurement data that
you want to preserve.
Remote
Command
:SYSTem:PUP:PROCess
Example
:SYST:PUP:PROC
After sending this command, you must wait for the instrument software to restart
Notes You cannot use *WAI or *OPC? to synchronize operation after a restart. This command stops and
restarts the instrument application, so the SCPI operation is terminated and restarted
A remote program must wait a fixed time before resuming sending commands to the instrument. The
appropriate wait time depends on which applications are pre-loaded