3 Short-Range Comms & IoT Mode
3.3 LoRa (CSS) Demodulation Measurement
Restart
The Restart function restarts the current sweep, or measurement, or set of
averaged/held sweeps or measurements. If you are Paused, pressing Restart does a
Resume.
The front-panel Restart key performs this exact same function
The Restart function is accessed in several ways:
Pressing the Restart key
Sending the remote command INIT:IMMediate
Sending the remote command INIT:RESTart
See "More Information" on page 423
Remote
Command
:INITiate[:IMMediate]
:INITiate:RESTart
Example
:INIT:IMM
:INIT:REST
Notes :INITiate:RESTart and :INITiate:IMMediate perform exactly the same function
Couplings Resets average/hold count k. The first sweep overwrites all active (update=on) traces with new current
data. For application modes, it resets other parameters as required by the measurement
Status Bits/OPC
dependencies
This is an Overlapped command
The STATus:OPERation register bits 0 through 8 are cleared
The STATus:QUEStionable register bit 9 (INTegrity sum) is cleared
The SWEEPING bit is set
The MEASURING bit is set
Backwards
Compatibility
Notes
For Spectrum Analysis mode in ESA and PSA, the Restart hardkey and the INITiate:RESTart command
restart trace averages (displayed average count reset to 1) for a trace in Clear Write, but did not
restart Max Hold and Min Hold
In the X-Series, the Restart hardkey and the INITiate:RESTart command restart not only Trace
Average, but Max Hold and Min Hold traces as well
For wireless comms modes in ESA and PSA, the Restart hardkey and the INITiate:RESTart command
restart every measurement, which includes all traces and numeric results. There is no change to this
operation
More Information
The Restart function first aborts the current sweep or measurement as quickly as
possible. It then resets the sweep and trigger systems, sets up the measurement
and initiates a new data measurement sequence with a new data acquisition
Short Range Comms & IoT Mode User's &Programmer's Reference 423