8 Trigger
8.1 Trigger
Preset 17
State Saved Saved in instrument state
Min 1
The minimum value is the minimum line and rolls over to the maximum value. The minimum line
number depends on which Field and standard are selected
Max The maximum value is the maximum line and rolls over to the minimum value. The maximum line
number depends on which Field and standard are selected
Field 1 (ODD):
–
Maximum line is 263 for formats NTSC-M, NTSC-Japan, NTSC-4.43, PAL-M and PAL-60
–
Maximum line is 313 for formats PAL-B, D, G, H, I, PAL-N, PAL-N Combin, and SECAM-L
Field 2 (EVEN):
–
The maximum line 262 for formats NTSC-M, NTSC-Japan, NTSC-4.43, PAL-M and PAL-60
–
The maximum line is 312 for formats PAL-B, D, G, H, I, PAL-N, PAL-N Combin, and SECAM-L
Field = ENTire Frame:
–
525, for formats NTSC-M, NTSC-Japan, NTSC-4.43, PAL-M and PAL-60
–
625, for formats PAL-B, D, G, H, I, PAL-N, PAL-N Combin, and SECAM-L
8.1.14 Field
Selects the Field on which to trigger:
Entire
Frame
ENTire
Causes the selected line number to be viewed as an offset into the
entire frame starting with line 1, the first line in Field One
Field One
ODD
Causes the selected line number to be viewed as an offset into the first
field starting with Line 1, the first line in Field One
Field Two
EVEN
Causes the selected line number to be viewed as an offset into the
second field. If Line 1 is selected, it is the 264th line of the frame
(NTSC-M, NTSC-Japan, NTSC-4.43, PAL-M, PAL-60) or the 314th line
of the frame (PAL-B,D,G,H,I, PAL-N, PAL-N-Combin, SECAM-L)
Remote Command
:TRIGger[:SEQuence]:TV:FMODe ENTire | ODD | EVEN
:TRIGger[:SEQuence]:TV:FMODe?
Example
:TRIG:TV:FMOD ENT
:TRIG:TV:FMOD EVEN
:TRIG:TV:FMOD ODD
Dependencies Only available in the Swept SA measurement
Only appears when TV is selected as the Trigger Source
2355 Short Range Comms & IoT Mode User's &Programmer's Reference