3 Short-Range Comms & IoT Mode
3.12 Monitor Spectrum Measurement
Backwards
Compatibility
SCPI
[:SENSe]:AVERage[:STATe] ON | OFF | 1 | 0
[:SENSe]:AVERage[:STATe]?
Backwards
Compatibility
Notes
Previous to X-Series, Averaging (also sometimes known as trace averaging) was global to all traces,
that is, it was either on or off for all active traces. The legacy command [:SENSe]:AVERage
[:STATe] ON|OFF|1|0 was used to turn Averaging on or off
In X-Series, Averaging is turned on or off on a per-trace basis, so it can be on for one trace and off for
another
For backwards compatibility, the old global Average State variable is retained solely as a legacy
variable, turned on and off and queried by the legacy command [:SENSe]:AVERage[:STATe]
OFF|ON|0|1. When Average is turned on, any trace in Clear/Write will get put into Average. While
Average is on, any trace put into Clear/Write by the old :TRAC:MODE command will instead get put
into Average. When Average is turned off, any trace in Average will get put into Clear/Write
Trace Type Details
Clear/Write
Each trace update replaces the old data in the trace with new data.
Pressing Clear/Write for the selected trace, or sending :TRAC:TYPE WRIT for the
specified trace, sets the trace type to Clear/Write and clears the trace, even if you
are already in Clear/Write. Then a new sweep is initiated. Trigger conditions must be
met before the sweep actually starts, and if in Single the sweep won’t start until
Restart is pressed.
Pressing Clear/Write stops the current sweep and initiates a new one, so Trace
Average, Max Hold and Min Hold data may be interrupted in mid-sweep when
Clear/Write is pressed, and therefore may not accurately reflect the displayed
count. Therefore, when Clear/Write is pressed for one trace, Trace Average, Max
Hold and Min Hold must restart for all traces.
When in Clear/Write, if a measurement-related instrument setting is changed (that
is, one which requires new data to be taken, like Center Frequency or Attenuation), a
new sweep is initiated but the trace is not cleared.
Trace Average
The instrument maintains and displays an average trace, which represents the
cumulative average on a point-by-point basis of the new trace data and previous
averaged trace data.
Pressing Trace Average (for the selected trace), or sending :TRAC:TYPE AVER (for
the specified trace), sets the trace type to Trace Average, clears the trace, initiates a
new sweep, and restarts the Average sequence.
1734 Short Range Comms & IoT Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals