4 System
4.6 Alignments
Query Time since Last Align VXT Transceiver
Remote Command
:CALibration:TIME:INTernal:VXT:TRANsceiver?
Example
:CAL:TIME:INT:VXT:TRAN?
Notes Value in hours since the last successful Align VXT Transceiver
Returns NaN if Align VXT Transceiver has never been performed on the instrument
Dependencies Only appears in M9410E/11E/15E/16E
State Saved No
Query Temperature of Last Align VXT Transceiver
Remote Command
:CALibration:TEMPerature:INTernal:VXT:TRANsceiver?
Example
:CAL:TEMP:INT:VXT:TRAN?
Notes Value in degrees Centigrade when the last successful Align VXT Transceiver was executed
Returns 9.91E+37(NaN) if Align VXT Transceiver has never been performed on the instrument
Dependencies Only appears in M9410E/11E/15E/16E
State Saved No
Query Time since Last Align Up Down Converter
Remote Command
:CALibration:TIME:UPDown:CONVerter?
Example
:CAL:TIME:UPD:CONV?
Notes Value in hours since the last successful Align Up Down Converter
Returns NaN if Align Up Down Converter has never been performed on the instrument
Dependencies Only appears in M9410E/11E/15E/16E
State Saved No
Query Temperature of Last Align Up Down Converter
Remote Command
:CALibration:TEMPerature:UPDown:CONVerter?
Example
:CAL:TEMP:UPD:CONV?
Notes Value in degrees Centigrade when the last successful Align Up Down Converter was executed
Returns 9.91E+37(NaN) if Align Up Down Converter has never been performed on the instrument
Dependencies Only appears in VXT models M9410A/11A and M9410E/11E
State Saved No
Query Time since Last Path Delay Calibration
Remote
Command
:CALibration:TIME:PDELay:SOURce?
Short Range Comms & IoT Mode User's &Programmer's Reference 1840