4.6 Alignments
Query Time since Last Align VXT Transceiver
Remote Command
:CALibration:TIME:INTernal:VXT:TRANsceiver?
Example
:CAL:TIME:INT:VXT:TRAN?
Notes Value in hours since the last successful Align VXT Transceiver
Returns NaN if Align VXT Transceiver has never been performed on the instrument
Dependencies Only appears in M9410E/11E/15E/16E
State Saved No
Query Temperature of Last Align VXT Transceiver
Remote Command
:CALibration:TEMPerature:INTernal:VXT:TRANsceiver?
Example
:CAL:TEMP:INT:VXT:TRAN?
Notes Value in degrees Centigrade when the last successful Align VXT Transceiver was executed
Returns 9.91E+37(NaN) if Align VXT Transceiver has never been performed on the instrument
Dependencies Only appears in M9410E/11E/15E/16E
State Saved No
Query Time since Last Align Up Down Converter
Remote Command
:CALibration:TIME:UPDown:CONVerter?
Example
:CAL:TIME:UPD:CONV?
Notes Value in hours since the last successful Align Up Down Converter
Returns NaN if Align Up Down Converter has never been performed on the instrument
Dependencies Only appears in M9410E/11E/15E/16E
State Saved No
Query Temperature of Last Align Up Down Converter
Remote Command
:CALibration:TEMPerature:UPDown:CONVerter?
Example
:CAL:TEMP:UPD:CONV?
Notes Value in degrees Centigrade when the last successful Align Up Down Converter was executed
Returns 9.91E+37(NaN) if Align Up Down Converter has never been performed on the instrument
Dependencies Only appears in VXT models M9410A/11A and M9410E/11E
State Saved No
Query Time since Last Path Delay Calibration
Remote
Command
:CALibration:TIME:PDELay:SOURce?
Short Range Comms & IoT Mode User's &Programmer's Reference 1840

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals