5 Preset
5.2 Mode Preset
Backwards
Compatibility
Notes
In X-Series, the legacy “Factory Preset” has been replaced by Mode Preset, which only presets the
currently active Mode, not the entire instrument. In X-Series, you preset the entire instrument by
using System, Restore System Defaults All, which behaves essentially the same way as restore System
Defaults did in ESA and PSA
There is also no “Preset Type” as there was in PSA. The green Mode Preset front-panel key does a
Mode Preset, and the User Preset front-panel key does a User Preset. The old PRESet:TYPE
command is ignored (without generating an error), and SYST:PRES without a parameter does Mode
Preset
The settings and correction data under the Input/Output front-panel key (examples: Input Z Corr, Ext
Amp Gain, etc.) are no longer part of any Mode, so they are not preset by Mode Preset. They are preset
by Restore Input/Output Defaults, Restore System Defaults All. Note that because "User Preset" on
page 1907 performs Recall State, and all these settings are saved in State, they are recalled when
using User Preset
1903 Short Range Comms & IoT Mode User's &Programmer's Reference