7 Save/Recall/Print
7.3 Save
–
Sweep Time
–
Sweep Time Auto
–
TriggerSource
–
Video Bandwidth
–
Y Axis Unit
Following the header entries above is a line containing only MeasResult1 and
MeasResult2, which flags the start of the measurement results. Each subsequent
line consists of two comma-separated values, the MeasResult1 value and the
MeasResult2 value.
–
MeasResult1 contains the same results as :MEAS|:READ|:FETCh:CHPower1
–
MeasResult2 contains the same results as :MEAS|:READ|:FETCh:CHPower2
The exported file is in CSV format. When imported into Microsoft Excel or a similar
spreadsheet application, a typical file appears as follows:
MeasResult
SA:CHP
A.10.53 N9030A
526 ALV ATP B1X B1Y B25 B40 BBA CR3 CRP DCF DDA DP2 DRD EA3 EDP
EMC EP1 ERC ESC ESP EXM FSA LFE LNP MAT MPB NFE NUL P26 PFR PNC
RTL RTS S40 SB1 SEC SM1 TVT YAS YAV
1
Auto Sweep Time Rules Normal
Average Mode Exponential
Average Number 10
Average State TRUE
Center Frequency 13255000000
Detector Average
IFGain FALSE
IFGainAuto FALSE
Impedance 50
Integ BW 2000000
Internal Preamp FALSE
Internal Preamp Band Low
PSD Unit DbmHz
Resolution Band Width 27000
Resolution Bandwidth Shape Gaussian
Short Range Comms & IoT Mode User's &Programmer's Reference 2269