3 Short-Range Comms & IoT Mode
3.4 HRP UWB Demodulation Measurement
knob, and it behaves as it normally would in manual mode
The Auto/Man state of (Mech) Atten is saved
The Auto/Man toggle on the (Mech) Atten control disappears, and the auto rules
are disabled
The Electronic Attenuator is set to 10dB less than the previous value of the
Mechanical Attenuator, within the limitation that it must stay within the range of
0 to 24dB of attenuation
Examples in the Dual-Attenuator configuration:
Mech Atten at 20dB. Elec Atten enabled, Mech Atten set to 10dB, and Elec
Atten set to 10dB. New total attenuation equals the value before Elec Atten
enabled
Mech Atten at 0dB. Elec Atten enabled, Mech Atten set to 10dB, and Elec Atten
set to 0dB. New total attenuation does not equal the value before Elec Atten
enabled
Mech Atten at 40dB. Elec Atten enabled, Mech Atten set to 10dB, and Elec
Atten set to 24dB. New total attenuation does not equal the value before Elec
Atten enabled
When the Electronic Attenuation is disabled from an enabled state:
The Elec Atten control is grayed out
The Auto/Man state of (Mech) Atten is restored
If now in Auto, (Mech) Atten recouples
If now in Man, (Mech) Atten is set to the value of total attenuation that existed
before the Elec Atten was disabled. The resulting value is rounded up to the
smallest value possible given the (Mech) Atten Step setting - (That is, 57dB
changes to 58dB when (Mech) Atten Step is 2dB)
Using the Electronic Attenuator: Pros and Cons
The electronic attenuator offers finer steps than the mechanical attenuator, has no
acoustical noise, is faster, and is less subject to wear.
The “finer steps” advantage of the electronic attenuator is beneficial in optimizing
the alignment of the instrument dynamic range to the signal power in the front
panel as well as remote use. Thus, you can achieve improved relative signal
measurement accuracy. Compared to a mechanical attenuator with 2dB steps, the
1dB resolution of the electronic attenuator only gives better resolution when the
odd-decibel steps are used. Those odd-decibel steps are less accurately calibrated
Short Range Comms & IoT Mode User's &Programmer's Reference 449

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals