3 Short-Range Comms & IoT Mode
3.5 Channel Power Measurement
Option # Description
"Best Close-in"
on page 641
4 or 1*
In instruments with EP0, emphasizes close-in phase noise
performance without regard to spur avoidance
In instruments without EP0, this setting is accepted but no action is
taken
"Best Spurs"
on page 642
5
In instruments with EP0, emphasizes spur avoidance over close-in
phase noise performance
In instruments without EP0, this setting is accepted but no action
taken
Auto - Automatically selects LO (local oscillator) phase noise behavior to
optimize dynamic range and speed for various instrument operating
conditions
*Dependent on Option EP0 installation. See "Best Close-in" on page 641 below.
The actual behavior varies somewhat depending on model number and option; for
example, you always get Fast Tuning by choosing Option #3, but in some models,
"Fast Tuning" on page 642 is identical in effect to "Best Close-in" on page 641.
Best Close-in
Without option EP0
:FREQ:SYNT 1
The LO phase noise is optimized for smaller offsets from the carrier, at the expense
of phase noise farther out.
The actual frequency offset within which noise is optimized is shown with in square
brackets, as this can vary depending on the hardware set in use. For example, in
some instruments this annotation appears as [offset <20 kHz]
With option EP0
:FREQ:SYNT 4
In instruments with Option EP0, the LO is configured for the best possible close-in
phase noise (offsets up to 600 kHz from the carrier), regardless of spurious products
that occur with some center frequencies. Because this is generally less desirable for
close-in measurements than the "Balanced" on page 641 setting, parameter 1
selects "Balanced" on page 641 in EP0 instruments, in the interests of optimizing
code compatibility across the family. Parameter 4 selects "Best Close-in" on page
641, which is usually not as good a choice as "Balanced" on page 641.
Balanced
:FREQ:SYNT 1
In instruments with EP0, the LO is configured for the best possible phase noise at
offsets up to 600 kHz from the carrier whenever there are no significant spurs within
Short Range Comms & IoT Mode User's &Programmer's Reference 641

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)