:SBUS<n> Commands 31
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1005
:SBUS<n>:MANChester:DSIZe
(see page 1610)
Command Syntax
:SBUS<n>:MANChester:DSIZe {AUTO | <#words>}
<#words> ::= from 1-255, in NR1 format
When the display format is WORD (see :SBUS<n>:MANChester:DISPlay), the
:SBUS<n>:MANChester:DSIZe command specifies the number of words in the
data field of your Manchester protocol definition.
AUTO is available as a selection only when the trailer field size is 0 (see
:SBUS<n>:MANChester:TSIZe).
Query Syntax
:SBUS<n>:MANChester:DSIZe?
The :SBUS<n>:MANChester:DSIZe? query returns the number of data field words
setting.
Return Format
<#words><NL>
<#words> ::= from 0-255, in NR1 format
In AUTO mode, the query returns 0.
In BIT format the only legal value is 0 (for AUTO).
See Also ":SBUS<n>:MANChester:DISPlay" on page 1004
":SBUS<n>:MANChester:HSIZe" on page 1006
":SBUS<n>:MANChester:SSIZe" on page 1010
":SBUS<n>:MANChester:TSIZe" on page 1016
":SBUS<n>:MANChester:WSIZe" on page 1017