1392 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
35 :TRIGger Commands
:TRIGger:ZONE<n>:STATe
(see page 1610)
Command Syntax
:TRIGger:ZONE<n>:STATe <on_off>
<n> ::= {1 | 2}
<on_off> ::= {{0 | OFF} | {1 | ON}}
<n> ::= 1-2 in NR1 format
The :TRIGger:ZONE<n>:STATe command sets the state for Zone 1 or Zone 2.
When a zone's state is on, and the Zone Trigger feature is on (see
":TRIGger:ZONE:STATe" on page 1388), that zone is actively being used to
qualify the trigger if it is not invalid or off-screen (see
":TRIGger:ZONE<n>:VALidity" on page 1391).
When the Zone Trigger feature is off, no zones are being used to qualify the
trigger, regardless of their individual states.
Note that :TRIGger:ZONE:STATe mimics the behavior of the [Zone] key on the front
panel, and :TRIGger:ZONE<n>:STATe mimics the behavior of the Zone 1 On and
Zone 2 On softkeys. At least one zone's state must be on for the Zone Trigger
feature (:TRIGger:ZONE:STATe) to be on. When the states of both individual zones
are turned off, Zone Trigger is automatically turned off. In this case, when Zone
Trigger is turned back on Zone 1 is forced to on. Otherwise, if at least one zone
was on when Zone Trigger was turned off, the same configuration of individual
zone on/off states will be restored when Zone Trigger is turned back on.
Query Syntax
:TRIGger:ZONE<n>:STATe?
The :TRIGger:ZONE<n>:STATe? query returns the state of Zone 1 or Zone 2.
Return Format
<on_off><NL>
<on_off> ::= {0 | 1}
See Also ":TRIGger:ZONE:STATe" on page 1388
":TRIGger:ZONE<n>:VALidity" on page 1391