:RECall Commands 29
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 819
:RECall:WMEMory<r>[:STARt]
(see page 1610)
Command Syntax
:RECall:WMEMory<r>[:STARt] [<file_name> | <data>]
<r> ::= 1 to (# ref waveforms) in NR1 format
<file_name> ::= quoted ASCII string
<data> ::= binary block data in IEEE 488.2 # format
The :RECall:WMEMory<r>[:STARt] command recalls a reference waveform.
The <data> option lets you recall a reference waveform from a local file on the
controller PC (instead of from a USB storage device connected to the
oscilloscope). In this case, your remote program reads data from the local ".h5"
format reference waveform file in the same way that setup files are restored to the
oscilloscope (see ":SYSTem:SETup" on page 1277).
See Also • "Introduction to :RECall Commands" on page 810
• ":RECall:FILename" on page 814
• ":SAVE:WMEMory[:STARt]" on page 851
NOTE
If a file extension is provided as part of a specified <file_name>, it must be ".h5".

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals