EasyManuals Logo

Keysight Technologies InfiniiVision 4000 X Series User Manual

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #383 background imageLoading...
Page #383 background image
:COUNter Commands 13
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 383
:COUNter:NDIGits
(see page 1610)
Command Syntax
:COUNter:NDIGits <value>
<value> ::= 5 normally, 8 with 10 MHz reference signal, in NR1 format
The :COUNter:NDIGits command sets the number of digits of resolution used for
the frequency or period counter.
The number of digits is not adjustable and is normally 5 digits. However, the
resolution can be increased to 8 digits by inputting an external 10 MHz reference
signal to the oscilloscope (on the rear panel 10MHz REF connector).
Higher resolutions require longer gate times, which cause the measurement times
to be longer as well.
Query Syntax
:COUNter:NDIGits?
The :COUNter:NDIGits? query returns the currently set number of digits of
resolution.
Return Format
<value><NL>
<value> ::= 5 or 8 in NR1 format
See Also • ":COUNter:CURRent" on page 380
• ":COUNter:ENABle" on page 381
• ":COUNter:MODE" on page 382
• ":COUNter:SOURce" on page 384

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals