772 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
28 :POWer Commands
:POWer:PSRR:WGEN:VOLTage
(see page 1610)
Command Syntax
:POWer:PSRR:WGEN:VOLTage <amplitude>[,<range>]
<amplitude> ::= amplitude in volts in NR3 format
<range> ::= {F20HZ | F100HZ | F1KHZ | F10KHZ | F100KHZ | F1MHZ
| F10MHZ | F20MHZ}
The :POWer:PSRR:WGEN:VOLTage command sets the waveform generator output
amplitude(s).
When the waveform generator amplitude profile is enabled (with the
:POWer:PSRR:WGEN:VOLTage:PROFile command), you can set an initial ramp
amplitude for each frequency range.
Without the <range> parameter, this command sets the waveform generator
output amplitude used when the amplitude profile is disabled.
Query Syntax
:POWer:PSRR:WGEN:VOLTage? [<range>]
The :POWer:PSRR:WGEN:VOLTage? query returns the waveform generator output
amplitude setting(s).
Return Format
<amplitude><NL>
<amplitude> ::= amplitude in volts in NR3 format
See Also • ":POWer:PSRR" on page 760
• ":POWer:PSRR:APPLy" on page 761
• ":POWer:PSRR:DATA" on page 762
• ":POWer:PSRR:FREQuency:MAXimum" on page 763
• ":POWer:PSRR:FREQuency:MINimum" on page 764
• ":POWer:PSRR:FREQuency:MODE" on page 765
• ":POWer:PSRR:PPDecade" on page 767
• ":POWer:PSRR:SOURce:INPut" on page 768
• ":POWer:PSRR:SOURce:OUTPut" on page 769
• ":POWer:PSRR:WGEN:LOAD" on page 771
• ":POWer:PSRR:WGEN:VOLTage:PROFile" on page 773