:POWer Commands 28
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 791
:POWer:SIGNals:VSTeady:ONOFf:ON
(see page 1610)
Command Syntax
:POWer:SIGNals:VSTeady:ONOFf:ON <value>[suffix]
<value> ::= Expected steady state output Voltage value in NR3 format
[suffix] ::= {V | mV}
The :POWer:SIGNals:VSTeady:ONOFf:ON command specifies the expected steady
state output DC voltage of the power supply for turn on analysis.
Query Syntax
:POWer:SIGNals:VSTeady:ONOFf:ON?
The :POWer:SIGNals:VSTeady:ONOFf:ON query returns the expected steady state
voltage setting.
Return Format
<value><NL>
<value> ::= Expected steady state output Voltage value in NR3 format
See Also • ":POWer:ONOFf:APPLy" on page 754
• ":POWer:SIGNals:AUTosetup" on page 776
• ":POWer:SIGNals:DURation:ONOFf:ON" on page 782
• ":POWer:SIGNals:IEXPected" on page 785
• ":POWer:SIGNals:OVERshoot" on page 786
• ":POWer:SIGNals:VMAXimum:ONOFf:ON" on page 789
• ":POWer:SIGNals:SOURce:CURRent<i>" on page 793
• ":POWer:SIGNals:SOURce:VOLTage<i>" on page 794

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals