:MTESt Commands 26
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 679
:MTESt:COUNt:WAVeforms
(see page 1610)
Query Syntax
:MTESt:COUNt:WAVeforms?
The :MTESt:COUNt:WAVeforms? query returns the total number of waveforms
acquired in the current mask test run.
Return Format
<count><NL>
<count> ::= number of waveforms in NR1 format.
See Also "Introduction to :MTESt Commands" on page 667
":MTESt:COUNt:FWAVeforms" on page 676
":MTESt:COUNt:TIME" on page 678
":MTESt:COUNt:RESet" on page 677
Example Code "Example Code" on page 667