:MEASure Commands 24
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 593
:MEASure:FFT:OBW
(see page 1610)
Command Syntax
:MEASure:FFT:OBW <percentage>[,<source>]
<percentage> ::= percent of spectral power occupied bandwidth is
measured for (in NR3 format)
<source> ::= {FUNCtion<m> | MATH<m> | FFT} (must be an FFT waveform)
<m> ::= 1 to (# math functions) in NR1 format
The :MEASure:FFT:OBW command installs an FFT analysis Occupied Bandwidth
measurement on screen.
Occupied Bandwidth measures the bandwidth (frequency range) containing some
percent (usually 99%) of the total spectral power. While 99% is the industry norm,
you can specify the percent you want to use in the measurement.
The center frequency used in the measurement is the one defined for the FFT
function, and the FFT span represents the total spectral power.
When this measurement is tracked with cursors, the cursors show the measured
bandwidth (frequency range).
Query Syntax
:MEASure:FFT:OBW? <percentage>[,<source>]
The :MEASure:FFT:OBW? query returns the measured Occupied Bandwidth value.
Return Format
<return_value><NL>
<return_value> ::= occupied bandwidth, Hz in NR3 format
See Also • ":MEASure:FFT:ACPR" on page 591
• ":MEASure:FFT:CPOWer" on page 592
• ":MEASure:FFT:THD" on page 594

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals