934 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:FLEXray:TRIGger
(see page 1610)
Command Syntax
:SBUS<n>:FLEXray:TRIGger <condition>
<condition> ::= {FRAMe | ERRor | EVENt}
The :SBUS<n>:FLEXray:TRIGger:TRIGger command sets the FLEXray trigger on
condition:
FRAMe — triggers on specified frames (without errors).
ERRor — triggers on selected active error frames and unknown bus conditions.
EVENt — triggers on specified FlexRay event/symbol.
Query Syntax
:SBUS<n>:FLEXray:TRIGger?
The :SBUS<n>:FLEXray:TRIGger? query returns the current FLEXray trigger on
condition.
Return Format
<condition><NL>
<condition> ::= {FRAM | ERR | EVEN}
See Also "Introduction to :TRIGger Commands" on page 1295
":TRIGger:MODE" on page 1308
":SBUS<n>:FLEXray:TRIGger:ERRor:TYPE" on page 935
":SBUS<n>:FLEXray:TRIGger:EVENt:AUToset" on page 936
":SBUS<n>:FLEXray:TRIGger:EVENt:BSS:ID" on page 937
":SBUS<n>:FLEXray:TRIGger:EVENt:TYPE" on page 938
":SBUS<n>:FLEXray:TRIGger:FRAMe:CCBase" on page 939
":SBUS<n>:FLEXray:TRIGger:FRAMe:CCRepetition" on page 940
":SBUS<n>:FLEXray:TRIGger:FRAMe:ID" on page 941
":SBUS<n>:FLEXray:TRIGger:FRAMe:TYPE" on page 942

Table of Contents