338 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
10 :CALibrate Commands
:CALibrate:STARt
(see page 1610)
Command Syntax
:CALibrate:STARt
The CALibrate:STARt command starts the user calibration procedure.
See Also • "Introduction to :CALibrate Commands" on page 332
• ":CALibrate:PROTected" on page 337
NOTE
Before starting the user calibration procedure, you must set the rear panel CALIBRATION
switch to UNPROTECTED, and you must connect BNC cables from the TRIG OUT connector to
the analog channel inputs. See the User's Guide for details.