:TRIGger Commands 35
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1383
:TRIGger:TV:UDTV:HSYNc
(see page 1610)
Command Syntax
:TRIGger:TV:UDTV:HSYNc {{0 | OFF} | {1 | ON}}
The :TRIGger:TV:UDTV:HSYNc command enables or disables the horizontal sync
control in the Generic video trigger.
For interleaved video, enabling the HSYNc control and setting the HTIMe
adjustment to the sync time of the probed video signal allows the ENUMber
function to count only lines and not double count during equalization.
Additionally, the Field Holdoff can be adjusted so that the oscilloscope triggers
once per frame.
Similarly, for progressive video with a tri-level sync, enabling the HSYNc control
and setting the HTIMe adjustment to the sync time of the probed video signal
allows the ENUMber function to count only lines and not double count during
vertical sync.
Query Syntax
:TRIGger:TV:UDTV:HSYNc?
The :TRIGger:TV:UDTV:HSYNc query returns the horizontal sync control setting.
Return Format
{0 | 1}
See Also • ":TRIGger:TV:STANdard" on page 1381
• ":TRIGger:TV:UDTV:HTIMe" on page 1384
• ":TRIGger:TV:UDTV:ENUMber" on page 1382
• ":TRIGger:TV:UDTV:PGTHan" on page 1385

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals