:POWer Commands 28
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 731
:POWer:DESKew
(see page 1610)
Command Syntax
:POWer:DESKew
The :POWer:DESKew command launches the auto deskew process on the
oscilloscope.
Before sending this command:
1 Demagnetize and zero-adjust the current probe.
Refer to the current probe's documentation for instructions on how to do this.
2 Make connections to the U1880A deskew fixture as described in the
oscilloscope's connection dialog or in the Power Measurement Application
User's Guide.
3 Make sure the voltage probe and current probe channels are specified
appropriately using the :POWer:SIGNals:SOURce:VOLTage1 and
:POWer:SIGNals:SOURce:CURRent1 commands.
The deskew values are saved in the oscilloscope until a factory default or secure
erase is performed. The next time you run the Power Application, you can use the
saved deskew values or perform the deskew again.
Generally, you need to perform the deskew again when part of the test setup
changes (for example, a different probe, different oscilloscope channel, etc.) or
when the ambient temperature has changed.
See Also • ":POWer:SIGNals:SOURce:VOLTage<i>" on page 794
• ":POWer:SIGNals:SOURce:CURRent<i>" on page 793
NOTE
Use the lowest attenuation setting on the high voltage differential probes whenever possible
because the voltage levels on the deskew fixture are very small. Using a higher attenuation
setting may yield inaccurate skew values (and affect the measurements made) because the
noise level is magnified as well.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals