1304 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
35 :TRIGger Commands
:TRIGger:HOLDoff:RANDom
(see page 1610)
Command Syntax
:TRIGger:HOLDoff:RANDom {{0 | OFF} | {1 | ON}}
The :TRIGger:HOLDoff:RANDom command enables or disables the random trigger
holdoff mode. This mode randomizes the holdoff time from one acquisition to the
next. The randomized holdoff time values will be between the values specified by
the :TRIGger:HOLDoff:MINimum and :TRIGger:HOLDoff:MAXimum commands.
The random trigger holdoff mode ensures that the oscilloscope re-arms after each
acquisition in a manner that minimizes or eliminates the likelihood of triggering at
the beginning of a DDR burst. Randomizing the holdoff time increases the
likelihood that the oscilloscope will trigger on different data phases of a
multi-phase (8 data transfer) burst. This mode mixes up the traffic pattern the
oscilloscope triggers on and is very effective when used on repeating patterns.
Query Syntax
:TRIGger:HOLDoff:RANDom?
The :TRIGger:HOLDoff:RANDom? query returns random trigger holdoff mode
setting.
Return Format
<setting><NL>
<setting> ::= {0 | 1}
See Also • ":TRIGger:HOLDoff:MAXimum" on page 1302
• ":TRIGger:HOLDoff:MINimum" on page 1303

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals