:SBUS<n> Commands 31
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1025
:SBUS<n>:NRZ:FSIZe
(see page 1610)
Command Syntax
:SBUS<n>:NRZ:FSIZe <#bits>
<#bits> ::= from 2-255, in NR1 format
When the NRZ bus display format (:SBUS<n>:NRZ:DISPlay) is BIT, the
:SBUS<n>:NRZ:FSIZe command lets you specify the total frame size of the NRZ
signal from 2 to 255 bits. This would be equivalent to the sum of the number of bits
in the header, data, and trailer fields in WORD format.
Query Syntax
:SBUS<n>:NRZ:FSIZe?
The :SBUS<n>:NRZ:FSIZe? query returns the specified total frame size.
Return Format
<#bits><NL>
See Also • ":SBUS<n>:NRZ:DISPlay" on page 1023
• ":SBUS<n>:NRZ:DSIZe" on page 1024
• ":SBUS<n>:NRZ:HSIZe" on page 1026
• ":SBUS<n>:NRZ:STARt" on page 1031
• ":SBUS<n>:NRZ:TSIZe" on page 1035
• ":SBUS<n>:NRZ:WSIZe" on page 1036